Healthsense Heart-Mate Blood Pressure Monitor BP-120 User Manual

cover-image

Content

Introduction of Heart-Mate Blood Pressure Monitor BP-120

The monitor features blood pressure measurement, alarm pulse rate measurement, talking function, and the result storage. The design provides you with one year of reliable service.
Readings taken by the BP-120 are equivalent to those obtained by a trained observer using the cuff and stethoscope auscultation method. This manual contains important safety and care information, and provides step-by-step instructions for using the product. Read the manual thoroughly before using the product.

Indications for Use

The HealthSense Blood Pressure Monitor is a digital monitor intended for use in measuring blood pressure and heartbeat rate with arm circumference ranging from 22 cm to 42 cm ( about 8¾?-16½? ).

It is intended for adult indoor use only.

LCD display signal

lcd_display_signal_healthsense_heart-mate_blood_pressure_monitor_bp-120.png
systolic_blood_pressure.png

 

Systolic blood pressure

 

High pressure result

diastolic_blood_pressure.png
Diastolic blood pressureLow pressure result
pulse_per_minute.png
Pulse per minuteBeats per minute, BPM
who_symbol.png
who_symbol.png
WHO SymbolShows BP level as per WHO standards
memory.png
MemoryThe displayed measurement values are from the memory.
low_battery.png
Low batteryBatteries are low and need to be replaced
irregular_heartbeat.png
Irregular heartbeatIrregular heartbeat Detection
current_time.png
Current TimeYear/Month/Day, Hour/Minute
heartbeat.png

 

HeartbeatHeartbeat detection during the measurement
user_a.png
User AStart measurement and save the measuring results for User A
user_b.png

 

User B

Start measurement and save the measuring results for User B
average_value_symbol.png
Average Value SymbolAverage reading will be displayed
body_movement.png
Body Movement Indicator

 

Detects body movement

Monitor Components

monitor_components_healthsense_heart-mate_blood_pressure_monitor_bp-120.png

List of Blood Pressure Monitor BP-120

list_healthsense_heart-mate_blood_pressure_monitor_bp-120.png
  1. Blood Pressure Monitor (Heart-Mate BP 120 Classic)
  2. Cuff (Type BF applied part) (22cm~42cm)
    (Please use a HEALTHSENSE-authorized cuff. For the size of the actual cuff please refer to the label on the attached cuff.)
  3. 4×AAA batteries
  4. User manual
  5. Warranty Card

BEFORE YOU START

The Choice of Power Supply

the_choice_of_power_supply_healthsense_heart-mate_blood_pressure_monitor_bp-120.png
  1. Battery-powered mode:  6VDC 4×AAA batteries
  2. AC adaptor powered mode: 6V 1A (Micro USB plug)
    (Please only use the recommended AC adaptor model). 
    Please unplug the adaptor to depart from the using utility power.

CAUTION

In order to get the best effect and protect your monitor,please use the right battery and special power adaptor which complies with CE safety standard.

Installing and Replacing the Batteries

installing_and_replacing_the_batteries_healthsense_heart-mate_blood_pressure_monitor_bp-120.png
  • Open the battery cover.
  • Install the batteries by matching the correct polarity, as shown.
  • Replace the cover.

Replace the batteries whenever the below happen

  • Thelow_battery.png shows
  • The display dims
  • The display does not light up

CAUTION

  • Remove batteries if the device is not likely to be used for some time.
    The old batteries are harmful to the environment, do not dispose with other daily trash.
  • Remove the old batteries from the device and follow your local recycling guidelines.
  • Do not dispose of batteries in fire. Batteries may explode or leak.

Measurement Principle

This product uses the Oscillometric Measuring method to detect blood pressure. Before every measurement, the unit establishes a “zero pressure” equivalent to the air pressure. Then it starts inflating the arm cuff, meanwhile, the unit detects pressure oscillations generated by beat-to-beat pulsatile, which is used to determine the systolic and diastolic pressure, and also pulse rate.

The device also compares the longest and the shortest time intervals of detected pulse waves to mean time interval then calculates standard deviation. The device will display a warning signal with the reading to indicate the detection of irregular heartbeat when the difference of the time intervals is over 25%.

Setting User, Date, Time & Measurement Unit

It is important to set the clock before using your blood pressure monitor, so that a time stamp can be assigned to each record that is stored in the memory. (The setting range of the year : 2016-2099 time format:12H/24H)

  1. When the monitor is off, press “ SET ” button about 8 seconds to enter YEAR setting mode. The YEAR will display.
  2. Press “MEM” button to change the [YEAR]. Each press will increase the numeral by one in a cycling manner.

    setting_user,_date,_time_&_measurement_unit_healthsense_heart-mate_blood_pressure_monitor_bp-120.png
  3. When you set the right year, press “ SET ” button to set down and turn to next step.
    Now you can set the [DATE] M/D.

    setting_user,_date,_time_&_measurement_unit_of_healthsense_heart-mate_blood_pressure_monitor_bp-120.png
  4. Repeat steps 2 and 3 to set the [MONTH] and [DAY].

    setting_user,_date,_time_&_measurement_unit_of_the_healthsense_heart-mate_blood_pressure_monitor_bp-120.png
  5. Repeat steps 2 and 3 to set the [TIME FORMAT] between 12H and 24H.
  6. Repeat steps 2 and 3 to set the [HOUR] and [MINUTE]. Shown below for 24H format

    setting_user,_date,_time_&_measurement_unit_of_the_healthsense-heart-mate_blood_pressure_monitor_bp-120.png
  7. Repeat steps 2 and 3 to set the measurement [UNIT] between mmHg and kPa.

    setting_user,_date,_time_&_measurement_unit_of_the_healthsense-heart-mate-blood_pressure_monitor_bp-120.png
  8. After setting Measurement Unit, press “ SET ” button to save the data and LCD will turn off.

    setting_user,_date,_time_&_measurement_unit_of_healthsense_heart-mate-blood_pressure_monitor_bp-120.png

Select the User

  1. When the monitor is off, press “ SET ” button once to enter user setting mode. The user ID will display.

    select_the_user_healthsense_heart-mate_blood_pressure_monitor_bp-120.png
  2. Press “ SET ” button to select the user ID between user A and user B.

    select_the_user_of_healthsense_heart-mate_blood_pressure_monitor_bp-120.png
  3. After selecting the suitable user ID, Press “ START/STOP ” button to confirm and turn off LCD.

MEASUREMENT

Tie the cuff

  1. Tie the cuff on your upper arm, then position the tube off-center toward the inner side of arm in line with the little finger.
  2. The cuff should be snug but not too tight. You should be able to insert one finger between the cuff and your arm.
  3. Sit comfortably with your tested arm resting on a flat surface.
  4. Patients with Hypertension:

    tie_the_cuff_healthsense_heart-mate_blood_pressure_monitor_bp-120.png
    • The middle of the cuff should be at the level of the right atrium of the heart; Before starting measurement, please sit comfortably with legs uncrossed, feet flat on the floor, back and arm supported. Rest for 5 minutes before measuring.
    • Wait at least 3 minutes between measurements. This allows your blood circulation to recover.
    • For a meaningful comparison, try to measure under similar conditions. For example, take daily measurements at approximately the same time, position of upper arm, or as directed by a physician.

Start the Measurement

  1. When the monitor is off, press the “ START/STOP ” button to turn on the monitor, and it will finish the whole measurement.(Take user A for example.) 

    start_the_measurement_healthsense_heart-mate_blood_pressure_monitor_bp-120.png
  2. Press the “ START/STOP ” button  to power off, otherwise it will turn off within 1 minute.
    Tips: Maximum 120 records are both for User A and User B.

Recall the Records

  1. When the monitor is off, press “MEM” button to show the average value of the latest three records. 
    (Note: If the records are less than 3 groups, the LCD will still display average record of the last saved memory)
  2. Press “MEM” button to get the record for your other stored memory (from M no.1 to M no. 120)
    (Note: The order, date and time of the record will be shown alternately)

    recall_the_records_healthsense_heart-mate_blood_pressure_monitor_bp-120.png
  3. If you want to check another user records, press “ START/STOP ” button to turn off the monitor. Now follow steps 1, 2 & 3 as per page 11 to select the user ID between user A and user B. After selecting user B repeat steps 1 & 2 as per page 14 to read average & memory readings for user B.

    recall_the_records_of_healthsense_heart-mate_blood_pressure_monitor_bp-120.png

CAUTION: The most recent record (1) is shown first. Each new measurement is assigned to the first (1) record. All other records are pushed back one digit (e.g., 2 becomes 3, and so on), and the last record (120) is dropped from the list.

DATA MANAGEMENT

Delete the Records

If you did not get the correct measurement, you can delete all results by following steps below. (Take User A for example.)

delete_the_records_healthsense_heart-mate_blood_pressure_monitor_bp-120.png
  1. Press “MEM” button to enter memory mode. The flash display will show average saved memory
  2. Press and hold “MEM” button to confirm deleting and the monitor will display as below
    Note: To exit out of delete mode without deleting any records, press “ START/STOP ” the button before pressing "MEM" to confirm any delete commands.
  3. If there is no record.
    The following display will show.

INFORMATION FOR USER

INFORMATION FOR USER

tips_for_measurement_healthsense_heart-mate_blood_pressure_monitor_bp-120.png
  1. Measurements may be inaccurate if taken in the following circumstances.
  2. Within 1 hour after dinner or drinking
  3. Immediate measurement after tea, coffee, smoking
  4. Within 20 minutes after taking a bath
  5. When talking or moving your fingers
  6. In a very cold environment
  7. When you want to discharge urine

SPECIFICATIONS

SpecificationDetails
Power supplyBattery powered mode: 6VDC (4×AAA batteries); AC adaptor powered mode: 6V = 1A (recommended model only)
Display modeDigital LCD V.A. 40 x 60 mm
Measurement modeOscillographic testing mode
Measurement rangeRated cuff pressure: 0kPa - 40kPa (0mmHg-300mmHg); Measurement pressure: 0-37.30kPa (0-280mmHg)
AccuracyPressure: ±0.4kPa (3mmHg); Pulse value: ±5%
Normal working conditionTemperature: 5°C to 40°C; Relative humidity: ≤85% RH; Atmospheric pressure: 86kPa to 106kPa
Storage & transportation conditionTemperature: -20°C to 60°C; Relative Humidity: 10% RH - 93% RH; Atmospheric Pressure: 50kPa - 106 kPa
Measurement perimeter of the upper armAbout 22cm - 42cm
Net WeightApprox. 225g (excluding dry cells and cuff)
External dimensionsApprox. 118 (W) x 110 (L) x 52 (H) mm
In the boxClassic BP-120, 4xAAA batteries, BP Cuff, User manual & Warranty card
Mode of operationContinuous operation
Degree of protectionType BF applied part
Protection against ingress of waterIP21
Software VersionV01

WARNING: No modification of this equipment is allowed.

Description

Its elegant, compact form makes the HealthSense Heart-Mate BP-120 easy to store and travel. The device's big LCD display displays measurements clearly. With an automated inflating mechanism and a comfortable cuff for most adult arm sizes. Hypertension indicators enable consumers to quickly see their blood pressure.
Multiple users can track their blood pressure with the monitor's memory feature, which remembers up to 90 readings per user. This function is important for trend tracking and healthcare provider data exchange.

Setup Guide

Steps to use the HealthSense Heart-Mate BP-120:

  1. Put the batteries in as directed.
  2. Wrap the cuff over your upper arm snugly but not too tightly.
  3. The measurement begins when you press start.
  4. It will automatically inflate the cuff and measure your blood pressure.
  5. Your pulse rate and findings will appear on screen.
  6. Select the right user profile before reading if many users share the device.

Maintenance

In order to get the best performance, please follow the instructions below

maintenance_.healthsense_heart-mate_blood_pressure_monitor_bp-120.png
  • Put in a dry place and avoid the sunshine
  • Avoid touching water, clean it with a dry cloth in case.
  • Avoid intense shaking and collisions
  • Avoid dusty and unstable temperature environment
  • Using wet cloths to remove dirt
  • Do not attempt to clean the reusable cuff with water and never immerse the cuff in water.

ABOUT BLOOD PRESSURE

What are systolic pressure and diastolic pressure? 

When ventricles contract and pump blood out of the heart, the blood pressure reaches its maximum value in the cycle, which is called systolic pressure. When the ventricles relax, the blood pressure reaches its minimum value in the cycle, which is called diastolic pressure.

what_are_systolic_healthsense_heart-mate_blood_pressure_monitor_bp-120.png

What is the standard blood pressure classification?

The blood pressure classification published by World Health Organization (WHO) and International Society of Hypertension (ISH) in 1999 is as follows:

CAUTION: Only a physician can tell your normal BP range.Please contact a physician if your measuring result falls out of the range. Please note that only a physician can tell whether your blood pressure value has reached a dangerous point.

what_is_the_standard_blood_pressure_healthsense_heart-mate_blood_pressure_monitor_bp-120.png

Level

Blood

Pressure (mm Hg)

 

Optimal

 

Normal

 

High-normal

 

Mild

 

Moderate

 

Severe

 

SYS

 

<120

 

120-129

 

130-139

 

140-159

 

160-179

≥180

 

DIA

 

<80

80-84

85-89

90-99

100-109

≥110

Irregular Heartbeat Detector
An irregular heartbeat is detected (irregular_heartbeat.png) when a heartbeat rhythm varies while the unit is measuring the systolic and diastolic blood pressure. During each measurement, 
this equipment records the heartbeat intervals and works out the standard deviation. If the calculated value is larger than or equal to 15,the irregular heartbeat symbol appears on the symbol when the measurement results are displayed.

CAUTION

The appearance of the IHB icon indicates that a pulse irregularity consistent with an irregular heart-beat was detected during measurement. Usually this is NOT a cause for concern. However, if the symbol appears often, we recommend you seek medical advice. Please note that the device does not replace a cardiac examination, but serves to detect pulse irregularities at an early stage.

Body movement Indicator

If you move during measurement, the movement error symbol (body_movement.png) will appear on the display. Keep still and repeat the measurement

body_movement_indicator_healthsense_heart-mate_blood_pressure_monitor_bp-120.png

Why does my blood pressure fluctuate throughout the day?

why_does_my_blood_pressure_healthsense_heart-mate_blood_pressure_monitor_bp-120.png
  1. Individual blood pressure varies multiple times everyday. It is also affected by the way you tie your cuff and your measurement position, so please take the measurement under the same conditions.
  2. If the person takes medicine, the pressure will vary more.
  3. Wait at least 3 minutes for another measurement.

Why do I get a different blood pressure at home compared to the hospital ?
The blood pressure is different even throughout the day due to weather,emotion, exercise etc, Also, there is the “white coat” effect, which means blood pressure usually increases in clinical settings.

is_the_result_the_same_healthsense_heart-mate_blood_pressure_monitor_bp-120.png
  • What you need to pay attention to when you measure your blood pressure at home:
  • If the cuff is tied properly.
  • If the cuff is too tight or too loose.
  • If the cuff is tied on the upper arm.
  • If you feel anxious.
  • Taking 2-3 deep breaths before beginning will be better for measuring.

Advice: Relax yourself for 4-5 minutes until you calm down.

Is the result the same if measuring on the right arm ?
It is ok for both arms, but there will be some different results for different people. We suggest you to measure on same arm every time.

How to Sit Correctly?

how_to_sit_correctly_healthsense_heart-mate_blood_pressure_monitor_bp-120.png

Safety Information

The signs below might be in the user manual, labeling or other component. They are the requirement of standard and using.

  • symbol.pngSymbol for “THE OPERATION GUIDE MUST BE READ”
  • “complies.pngSymbol for “COMPLIES WITH MDD 93/42/EEC REQUIREMENTS”
  • serial_number.pngSymbol for “SERIAL NUMBER”
  • “direct_current.pngSymbol for “DIRECT CURRENT”
  • “manufacture.pngSymbol for “MANUFACTURE DATE”
  • type_bf_applied.pngSymbol for “TYPE BF APPLIED PARTS”
  • “environment.pngSymbol for “ENVIRONMENT PROTECTION - Electrical waste products should not be disposed of with household waste. Please recycle where facilities exist. Check with your local authority or retailer for recycling advice”
  • caution.pngCaution: These notes must be observed to prevent any damage to the device.

CAUTION

  • The device is not suitable for measuring the blood pressure of children. Ask your doctor before using it on older children.
  • The device is not suitable for use on pregnant women, patients with implanted, electrocical devices, patients with pre-elcampsia, premature ventricular beats, atrial fibrillation, peripheral, arterial disease and patients undergoing intravascular therapy or arterio-venous shunt or people who received a mastectomy. Please consult your doctor prior to using the unit if you suffer from illnesses
  • Do not take any therapeutic measures on the basis of a self measurement. Never alter the dose of a medicine prescribed by a doctor. Consult your doctor if you have any question about your blood pressure.
  • Please keep the unit out of reach of infants, children or pets, since inhalation or swallowing of small parts is dangerous or even fatal.

CAUTION

  • This device is intended only for adult use in homes.
  • This device is intended for non-invasive measuring and monitoring of arterial blood pressure. It is not intended for use on extremities other than arm or for functions other than obtaining a blood pressure measurement.
  • If you experience discomfort during a measurement, such as pain in the arm or other complaints, press the START/STOP button to release the air immediately from the cuff. Loosen the cuff and remove it from your arm.
  • On the rare occasion of a fault causing the cuff to remain fully inflated during measurement, open the cuff immediately. Prolonged high pressure (cuff pressure >300mmHg or constant pressure 15mmHg for more than 3 minutes) applied to the arm may lead to an ecchymosis.
  • Too frequent and consecutive measurements could cause disturbances in blood circulation and injuries.
  • Do not wrap the cuff on the same arm which other monitoring ME EQUIPMENT is applied simultaneously, because this could cause temporary loss of function of those simultaneously-used monitoring ME EQUIPMENT.
  • Don’t kink the connection tube during use, otherwise, the cuff pressure may continuously increase which can prevent blood flow and result in harmful injury to the PATIENT.
  • The maximum temperature that the applied part can be achieved is 42.5? while the environmental temperature is 40? .(only for wrist blood pressure monitor)
  • The equipment is not AP/APG equipment and not suitable for use in the presence of a flammable anesthetic mixture with air of with oxygen or nitrous oxide.
  • This unit is not suitable for continuous monitoring during medical emergencies or operations. This device cannot be used with HF surgical equipment at the same time.
  • This device is not intended for patient transports outside a healthcare facility.
  • To avoid measurement errors, please avoid the condition of strong electromagnetic field radiated interference signal or electrical fast transient/burst signal.
  • The operator shall not touch output of batteries/adapter and the patient simultaneously. Manufacturer will make available on request circuit diagrams, component parts list etc.
  • The materials of the cuff have been tested and found to comply with requirements of ISO 10993-5:2009 and ISO 10993-10:2010. It will not cause any potential sensization or irritation reaction.Never apply the cuff over hurt skin.
  • Do not use the unit in case of existing polyester resp. synthetic allergies.
  • Be careful to strangulation due to cables and hoses, particularly due to excessive length.
  • Do not connect the air hose to other medical equipment, as this could cause air to be pumped into intravascular systems or high pressure, what could lead to dangerous injuries.
  • Before use, make sure the device functions safely and is in proper working condition.
  • Please use the device under the environment which was provided in the user manual. Otherwise, the performance and lifetime of the device will been impacted and reduced.
  • Please use ACCESSORIES and detachable parts specified/ authorised by MANUFACTURER. Otherwise, it may cause damage to the unit or danger to the user/patients.
  • Please dispose of ACCESSORIES, detachable parts, and the ME EQUIPMENT according to the local guidelines.
  • Please do not attempt to repair the unit yourself in the event of malfunctions. Only have repairs carried out by authorized service centers.
  • Please report to Manufacturer if any unexpected operation or events occur.
  • The device doesn’t need to be calibrated in two years of reliable service.
  • Please use the soft cloth to clean the whole unit. Don’t use any abrasive or volatile cleaners.

Troubleshooting

TROUBLESHOOTING

This section includes a list of error messages and frequently asked questions for problems you may encounter with your blood pressure monitor. If the products not operating as you think it should, check here before arranging for servicing.

SYMBOLNo display appears
Er 1 
Er 2 
Er 3 
Er 4 
Er 5 
Er 6 
Er 7 
CAUSEWim roger placement
CORRECTIONReplace both batteries with new ones. Check the battery installation of the battery polarities.
Sensor abnormalSend it to the local distributor for proper aid of the unit.
Measurement result isSend it to local distributor for processing.
Pipe blockedWrap the cuff properly to ensure air flow.
Heart rate area displayHeart rate is more than 200 beats/min.
Heart rate area displayHeart rate is less than 40 beats/min.

PROBLEM

No Power

CHECK THIS

Check the battery power. Check the battery polarity position

REMEDY

Replace with new sets of batteries. Proper installation of batteries polarities

No inflation

Check the air hose plug

Tightly insert plug in air socket

Err and stop working

Movement of arm or body while measurement

Talking while measurement

Keep the body peaceful Keep quiet when measuring

Cuff leak

Whether the cuff wrap is too loose. Whether the cuff is broken

Wrap the cuff tightly Change to new cuff

AUTHORIZED COMPONENTS

  1. Please use the HEALTHSENSE authorized adapter or adapter with below power specification (excluded).
  2. Please use HEALTHSENSE BP series cuff (22-42 cm)
authorized_components_healthsense_heart-mate_blood_pressure_monitor_bp-120.png

Warranty

HealthSense India warranty covers defects in material and workmanship of the product under normal use for a period of one (1) year from the date of retail purchase. The warranty does not cover damages resulting from misuse, abuse, immersion, normal wear and tear or unauthorized modification.

Complied European Standards List

Risk management

ISO/EN 14971:2012 Medical devices — Application of risk management to medical devices

Labeling

EN 980 Symbols for use in the labelling of medical devices

User manual

EN 1041: 2008 Medical equipment manufacturers to provide information

 

 

General Requirements for Safety

EN 60601-1: 2006+A1:2013 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance

IEC/EN 60601-1-11: 2010    Medical electrical equipment -- Part 1-11: General requirements for basic safety and essential performance - Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare

environment

 

Electromagnetic compatibility

IEC/EN 60601-1-2:2007 Medical electrical equipment

- Part 1-2: General requirements for basic safety and essential performance - Collateral standard:Electromagnetic compatibility - Requirements and tests

 

Performance requirements

EN ISO 81060-1:2012(new) Non-invasive sphygmomanometers - Part 1: Requirements and test methods for non-automated measurement type (ISO 81060-1:2007)

EN 1060-3:1997+A2:2009 Non-invasive blood

pressure Part 3: Supplementary requirements for electromechanical blood pressure measuring system

Clinical investigation

EN 1060-4: 2004 Automatic Blood Pressure Monitor overall system Interventional accuracy of the testing process

 

 

Usability

IEC/EN 60601-1-6: 2010   Medical electrical equipment -- Part 1-6: General requirements for basic

safety and essential performance - Collateral Standard: Usability

IEC/EN 62366: 2007 Medical devices - Application of usability engineering to medical devices

Software life-cycle

processes

IEC/EN 62304:2006+AC: 2008 Medical device software - Software life cycle processes

EMC Guidance

  1. This equipment needs to be installed and put into service in accordance with the information provided in the user manual;
  2. Wireless communications equipment such as wireless home network devices, mobile phones, cordless telephones and their base stations, walkie-talkies can affect this equipment and should be kept at least a distance d=3,3m away from the equipment.

(Note: As indicated in Table 6 of IEC 60601-1-2:2007 for ME EQUIPMENT, a typical cell phone with a maximum output power of 2 W yields d=3, 3m at an IMMUNITY LEVEL of 3V/m)

Heart-Mate Blood Pressure Monitor BP-120 Pros & Cons

Pros

  • Accurate and reliable blood pressure readings
  • Ease of use with automatic inflation and deflation
  • Large LCD display for clear visibility
  • Memory storage for multiple users
  • Affordable price point compared to other advanced monitors

Cons

  • Limited cuff size may not fit all arm sizes comfortably
  • No Bluetooth or Wi-Fi connectivity for data transfer
  • Battery life could be longer based on usage frequency

Customer Reviews

Customers have praised the HealthSense Heart-Mate BP-120 for its accuracy and ease of use. Many have found it to be a reliable tool for monitoring blood pressure at home. Some users have noted that the cuff could be more adjustable to fit different arm sizes comfortably. Overall, the device has received positive feedback for its value and performance.

Most Common Complaints

  • Some users have reported issues with the cuff size not being adjustable enough.
  • A few customers have mentioned that the battery life could be longer.

Faqs

How accurate is the Monitor?

The HealthSense is designed to provide highly accurate readings, meeting international standards for blood pressure monitors. It ensures reliable measurements of both systolic and diastolic blood pressure as well as pulse rate.

Can multiple users share the same device?

Monitor BP-120 supports multiple users. It has a memory function that can store up to 90 readings per user, allowing different individuals to track their blood pressure independently.

What is the recommended cuff size for this monitor?

The recommended cuff size for the HealthSense Heart-Mate BP-120 is 22-32 cm. This range is suitable for most adult arm sizes but may not fit all users comfortably.

How do I know if my reading is within a healthy range?

The device includes a hypertension indicator that provides a quick visual reference to help you understand your blood pressure levels. You can also refer to the guidelines provided in the user manual or consult with a healthcare professional.

Can I transfer my readings to my smartphone or computer?

The HealthSense Heart-Mate BP-120 does not have Bluetooth or Wi-Fi connectivity. However, you can manually record your readings and transfer them to your digital health records or share them with your healthcare provider.

What type of batteries does the device use?

BP-120 uses standard AA batteries. It is recommended to use high-quality batteries for optimal performance and longer battery life.

How often should I calibrate my blood pressure monitor?

It is recommended to have your blood pressure monitor calibrated every 6-12 months or as specified in the user manual. You can contact the manufacturer or a healthcare professional for calibration services.

Is this device suitable for pregnant women?

While the HealthSense Heart-Mate BP-120 is generally safe for use, pregnant women should consult with their healthcare provider before using any new medical device. This ensures that it meets their specific health needs and provides accurate readings during pregnancy.

What is included in the packaging?

The packaging includes the blood pressure monitor, cuff, batteries, user manual, and sometimes additional accessories like a storage case or quick start guide.

What kind of warranty does this product come with?

The HealthSense Heart-Mate BP-120 typically comes with a standard warranty from the manufacturer. Details about the warranty period and coverage can be found in the user manual or by contacting customer support.

Rascal Pet Trailer 13-SC315

RIVER 2 Pro Power Station

Leave a Comment

Name

Email

Comment