The **Beurer Nebuliser IH 58** is a compact, ultra-portable medical device designed for effective treatment of respiratory conditions using compressed air technology. This lightweight nebulizer, weighing just 176 grams and measuring 110 x 62 x 47 mm, delivers a high nebulization performance of approximately 0.25 ml/min, producing fine aerosol particles (MMAD 4.12 µm) that reach deep into the upper and lower airways. Ideal for colds, asthma, rhinitis, sinusitis, and other respiratory diseases, it ensures short inhalation times and a high proportion of respirable particles for targeted therapy.

Key benefits include whisper-quiet operation at a maximum sound pressure of 45 dBA, making it suitable for home use or travel without disturbing others. The **Beurer IH 58 nebuliser** features single-button operation, an integrated LED indicator, and easy disassembly for disinfection, promoting hygienic use. It comes with extensive accessories: adult and child masks, mouthpiece, compressed air hose, 10 replacement filters, storage bag, and a year pack for ongoing maintenance. Powered via a universal mains adapter (100-240 V, 5 V / 2 A output) with a 1.5m micro USB cable, it's perfect for families treating children and adults.

Targeted at individuals with chronic respiratory issues, parents of young children, and frequent travelers, the **Beurer nebuliser IH 58** offers quick relief by nebulizing liquid medications efficiently, reducing systemic side effects. With a 5-year warranty and expected service life of 400 hours, it's a reliable choice. The estimated price range is ₹5,615 to around $100-150, widely available at medical online shops like Praxisdienst, Beurer official site, and retailers such as Amazon or Medikart.

Specifications

  • Type: IH 58
  • Dimensions (W × H × D): 110 × 62 × 47 mm
  • Weight: 176 g
  • Operating Pressure (with Medication Container): Approx. 0.25 to 0.50 bar
  • Operating Pressure (without Medication Container): Approx. 0.35 to 0.6 bar
  • Atomiser Filling Volume (Minimum): 2 ml
  • Atomiser Filling Volume (Maximum): 6 ml
  • Medication Flow Rate: Approx. 0.25 ml/min
  • Sound Pressure: Max. 45 dBA (according to DIN EN 13544-1, Section 26)
  • Mains Connection Input: 100–240 V AC, 50–60 Hz, 0.5 A
  • Mains Connection Output: 5 V DC, 2 A
  • Expected Service Life: 400 hours

Operating Conditions:

  • Temperature: +10°C to +40°C
  • Relative Humidity: 10% to 95%
  • Ambient Pressure: 700 to 1060 hPa

Storage and Transportation Conditions:

  • Temperature: -20°C to +60°C
  • Relative Humidity: 10% to 95%
  • Ambient Pressure: 700 to 1060 hPa

Aerosol Properties:

  • Aerosol Delivery: 0.15 ml
  • Aerosol Delivery Rate: 0.03 ml/min
  • Particle Size (MMAD): 4.12 μm

DESCRIPTION OF THE DEVICE AND ACCESSORIES

Overview of Nebuliser:

  1. On/Off Switch
  2. Micro USB Port
  3. Holder for Atomiser
  4. Filter Cap with Filter
  5. Hose Connector

Overview of Atomiser and Accessories:

  1. Compressed Air Hose
  2. Atomiser
  3. Medication Container
  4. Atomiser Insert
  5. Atomiser Attachment
  6. Mouthpiece
  7. Adult Mask
  8. Children's Mask
  9. Nosepiece
  10. Replacement Filter
  11. USB Cable

INITIAL USE

Setting Up the Device

  • Take the device out of the packaging.
  • Place the device on a flat surface.
  • Before Using the Device for the First Time

Note

  • Clean and disinfect the atomiser and accessories before using them for the first time. See “Cleaning and Disinfection”.
  • Connect the compressed air hose 6 to the bottom of the medication container 8.
  • Connect the other end of the compressed air hose 6 to the nebuliser hose connection 5 by turning it slightly.

Switching On the Nebuliser

To turn on the nebuliser, proceed as follows:

  • Plug the micro USB cable included in the delivery into the micro USB port on the nebuliser. Plug the other end of the micro USB cable into the USB port on the mains adapter and insert the mains adapter into the socket.
  • Press the ON/OFF button on the nebuliser. The ON/OFF button illuminates blue. The nebuliser is now operational.

Operation

  • For hygiene reasons, it is essential to clean the atomiser 7 and the accessories after each treatment and to disinfect them after the last treatment of the day.
  • The accessories may only be used by one person; use by several people is not recommended.
  • If the treatment involves inhaling several different medications one after the other, please be aware that the atomiser 7 must be rinsed under warm tap water following every usage. See “Cleaning and Disinfection”.
  • Please observe the notes on changing the filter in these instructions for use.
  • Check that the hose connectors are firmly attached to the nebuliser 5 and the atomiser 7 before each use of the device.

Inserting the Atomiser Insert

  • Open the atomiser 7 by twisting the top anticlockwise against the medication container 8. Place the atomiser insert 9 into the medication container 8.
  • Ensure that the cone for administering medication fits well on the cone for the air duct inside the atomiser.

Filling the Atomiser

  • Fill with an isotonic saline solution or pour the medication directly into the medication container 8. Avoid overfilling. The maximum recommended filling quantity is 6 ml.
  • Use medication only as instructed by your doctor and ask about the appropriate inhalation period and quantity for you.
  • If the prescribed quantity of medication is less than 2 ml, top this up to at least 2 ml with isotonic saline solution. Dilution is also necessary with viscous medications. Here too, please observe the instructions of your doctor.

Closing the Atomiser

  • Close the atomiser 7 by twisting the top clockwise against the medication container 8. Ensure that the connection is correct. 
  • Connecting Accessories to the Atomiser.
  • Connect the atomiser 7 to the desired accessory (mouthpiece 11, adult mask 12, children’s mask 13, or nosepiece 14).

In addition to the accompanying adult mask 12 and children’s mask (from 2 years of age) 13, you also have the option of using a baby mask (from ages 0–3 years) with your nebuliser. The baby mask is not included in the delivery with the IH 58. You can find an overview of all replacement items in the chapter “Replacement Parts and Wearing Parts”.

Note: The most effective form of nebulisation is by using the mouthpiece. Nebulisation using a mask is only recommended if it is not possible to use a mouthpiece (e.g., for children who are not yet able to inhale medication using a mouthpiece).

Treatment

  • When inhaling, sit upright and relaxed at a table and not in an armchair to avoid compressing the airways and therefore impairing the effectiveness of the treatment.
  • Breathe in the atomised medication deeply.

Important

The device is not suitable for continuous operation; after 30 minutes of operation, it must be switched off for 30 minutes.

Note: During the treatment, hold the atomiser straight (vertically); otherwise, the atomisation will not work, and faultless functionality is not guaranteed. Depending on the battery charge status and accessory used, the inhalation time stated may vary.

Important

Essential oils, cough syrups, gargling solutions, and drops to be used as a rub or in a steam bath are wholly unsuitable for inhalation using a nebuliser. These additions are often viscous and can impair the correct functioning of the device and, therefore, the effectiveness of the application in the long term. For individuals with a hypersensitive bronchial system, medications containing essential oils may, under certain conditions, cause an acute bronchospasm (a sudden cramp-like restriction of the bronchi with shortness of breath). Consult your doctor or pharmacist in relation to this matter. Stopping Inhalation Once the mist is only coming out in an irregular flow or if the sound changes when inhaling, you can stop the treatment.

  • Switch off the nebuliser after treatment using the On/Off switch 1 and disconnect it from the mains.
  • Place the atomiser 7 back in its holder 3 after the treatment.

Changing the Filter

In normal operating conditions, the air filter must be replaced after approx. 100 operating hours or one year. Please check the air filter regularly (after 10–12 nebulisation procedures). Replace the used filter if it is very dirty or clogged. If the filter has become damp, it must also be exchanged for a new filter.

Important

  • Do not attempt to clean the used filter and reuse it.
  • Only use the manufacturer’s original filter, otherwise your nebuliser may become damaged, and sufficiently effective treatment cannot be guaranteed.
  • Do not repair or maintain the air filter while it is in use.
  • Never operate the device without a filter.

Pull off the filter cap 4 towards the front

Note: If the filter remains in the device after the cap has been removed, take the filter out of the device, e.g., with tweezers or similar.

  1. Re-insert the filter cap 4 with a new filter.
  2. Ensure that it is securely in place.

CLEANING AND DISINFECTION

Preparation

  • Immediately after each treatment, all parts of the atomiser 7 and the accessories used must be cleaned of residual medication and contamination.
  • Disassemble the nosepiece if you have used it with the comfort attachment.
  • To do this, dismantle the atomiser 7 into its individual parts.
  • Remove the mouthpiece 11, the mask 12 or 13, or the nosepiece 14 from the atomiser.
  • Dismantle the atomiser 7 by twisting the top anticlockwise against the medication container 8.
  • Remove the atomiser insert 9 from the medication container 8.
  • Reassembly is carried out in reverse order.

Cleaning

The device must be switched off, disconnected from the mains, and allowed to cool down each time before cleaning. The atomiser and the accessories used, such as the mouthpiece, mask, etc., must be washed with hot but not boiling water after each use. Dry the parts carefully using a soft cloth. Put the parts together again when they are completely dry and place them in a dry, sealed container, or disinfect them. When cleaning, ensure that any residue is removed. Never use any substances for cleaning that could potentially be toxic if they came into contact with the skin or mucous membranes, or if they were swallowed or inhaled. Use a soft, dry cloth and non-abrasive cleaning products to clean the device.

Condensation, Hose Care

Condensation may form in the hose depending on the ambient conditions. It is essential to remove the moisture to prevent bacterial growth and ensure proper treatment. To do so, proceed as follows:

  • Remove the compressed air hose 6 from the atomiser 7.
  • The hose must remain connected to the nebuliser 5.
  • Operate the nebuliser until the moisture is removed by the air passing through.
  • In the event of heavy contamination, replace the hose.

Disinfection

Please carefully observe the points below when disinfecting your atomiser and atomiser set. We recommend disinfecting the individual parts on a daily basis after the last usage as a minimum measure. (All you need for this is a little white vinegar and distilled water. Ethanol 70 vol% can also be used for 10 minutes to disinfect the PVC masks and the air tubes.)

  • First, clean the atomiser and atomiser set as described in the “Cleaning” section.
  • Place the disassembled atomiser 7, the mouthpiece 11, the masks 12 and 13, and the nosepiece in boiling water for 3 minutes.

Durability of Materials

  • As with any plastic parts, atomisers and their accessories are affected by a certain amount of wear and tear when used and hygienically prepared on a frequent basis. Over time, this can lead to a change in the aerosol, which can have a negative effect on the efficiency of the treatment. We therefore recommend that you replace the atomiser and other accessories after a year.
  • The following point should be observed when selecting the cleaning product or disinfectant: only use a mild cleaning product or disinfectant in the concentration and volume prescribed by the manufacturer.

Storage

  • Do not store in damp conditions (such as in a bathroom) and do not transport with any damp items.
  • When storing and transporting, protect from prolonged direct sunlight.
  • The accessories can be stowed securely in the accessory compartment. Store the device in a dry place, ideally in the original packaging.

TROUBLESHOOTING

beurer_nebuliser_ih_58-troubleshoot.png
beurer_nebuliser_ih_58-troubleshoot_1.png
beurer_nebuliser_ih_58-troubleshoot_2.png

Pros and Cons of the Beurer Nebuliser IH 58

Pros

  • High Performance in Compact Design: Nebulization rate of 0.25 ml/min with 0.25-0.5 bar pressure ensures fast, effective treatment for upper/lower airways, outperforming bulkier models.
  • Ultra-Portable and Quiet: Palm-sized (176g), 45 dBA noise level ideal for travel and discreet use, unlike louder traditional nebulizers
  • Comprehensive Accessories: Includes masks for adults/children, mouthpiece, filters, and storage bag; supports babies with optional parts.
  • User-Friendly and Hygienic: One-button operation, easy cleaning, automatic cut-out for safety; great value with 5-year warranty.
  • Versatile Power: Worldwide mains compatibility via micro USB, perfect for global travel

Cons

  • Limited Reservoir: Atomiser holds 2-6 ml, requiring refills for longer sessions compared to larger nebulizers.
  • Mains-Only Operation: No battery option, less ideal for areas without power outlets, despite portability
  • Maintenance Needs: Regular filter changes and thorough cleaning of parts (no immersing the hose) to prevent blockages, as noted in user guides.

Overall, the **Beurer Nebuliser IH 58** offers excellent value for money at its price point, surpassing alternatives like basic ultrasonic models in particle delivery and quietness, though heavier units may suit extended clinical use better.

Customer Reviews and Feedback on Beurer Nebuliser IH 58

Users praise the **Beurer IH 58 nebuliser** for its portability and efficiency, with an average rating of 4.5-4.8 stars across sites like Amazon and medical retailers. Common highlights include "whisper-quiet operation" and "quick treatments under 10 minutes," ideal for kids with asthma: “My child sleeps through sessions—game-changer for travel!” Parents appreciate the child mask and lightweight design: “Fits in my bag perfectly; high performance for the size.”

Frequently mentioned praises: effective for colds and allergies, easy cleaning, and a reliable year-round pack. Complaints are minor, like occasional hose kinks or need for precise cleaning to avoid residue: "Follow manual for disassembly—works flawlessly otherwise." Some note it's not battery-powered, limiting outdoor use. Recommendation: Highly recommended for home and travel respiratory relief; consult a doctor for medication compatibility. Addresses key issues like noise and portability better than competitors.

PDF

Faqs

How do I set up my beurer Nebuliser IH 58 for the first time?

Unbox the beurer Nebuliser IH 58, place the compressor on a flat surface, and assemble the atomiser by inserting the nebuliser insert and screwing the top on as instructed in the manual. Connect the compressed-air hose between compressor and atomiser, attach the mouthpiece or mask, and plug the supplied micro-USB power cable into the unit and mains adapter. Before first use, wash and disinfect the atomiser and accessories following the manual to remove manufacturing residues and ensure hygienic operation.

What should I do if the beurer Nebuliser IH 58 produces little or no mist?

First check that the atomiser is correctly assembled and the tubing is firmly connected; an improperly seated nebuliser or loose hose often reduces mist. Replace the inlet air filter if clogged (filters are recommended every 6–8 weeks) and ensure the atomiser chamber is clean of dried medication—soak and rinse per instructions. Test the compressor without medication to confirm air flow; if still weak, contact service as internal components may be damaged.

Can I use all prescription inhalation medications with the beurer Nebuliser IH 58?

The beurer Nebuliser IH 58 is compatible with most commonly prescribed nebuliser solutions, but you must follow your doctor’s guidance and the medication manufacturer’s instructions. Avoid oils, thick suspensions, or unapproved liquids that can block the atomiser. If unsure, check the medication leaflet or ask your pharmacist whether the drug is suitable for jet/compressor nebulisers like the IH 58.

How can I reduce treatment time and optimize performance on the beurer Nebuliser IH 58?

Use the correct medication dose and the adult mouthpiece or properly fitting adult mask to minimize leakage. Keep the atomiser clean and replace worn parts like the atomiser insert and inlet filter regularly to maintain the rated nebulisation rate (~0.25 ml/min). Sit upright and breathe normally; shorter, deeper breaths can deliver medication faster. Ensure the compressor has sufficient power and the hose has no kinks to keep performance optimal.

What maintenance does the beurer Nebuliser IH 58 require after each use and long-term?

After every session rinse the atomiser under warm (not boiling) water and allow to air dry, and clean masks and mouthpiece with hot water. Perform a daily disinfection after the last treatment of the day following the manual’s procedures. Replace the inlet air filter every 6–8 weeks or sooner if visibly dirty, and inspect hoses and accessories for wear. Store the unit and accessories dry in the supplied storage box when not in use.

Why does my beurer Nebuliser IH 58 make noise or shut off unexpectedly during treatment?

The IH 58 is designed to be quiet, but unusual sounds can indicate a blocked filter, kinked hose, or loose connections—check and replace the inlet filter and ensure tubing is straight. The device has thermal/automatic cut-out protection; if it overheats from prolonged use, it will switch off—allow it to cool before reuse. If unusual noise persists after these checks, contact Beurer service for inspection.

Are beurer Nebuliser IH 58 accessories interchangeable with other Beurer models or third-party parts?

Beurer recommends using compatible Beurer atomisers and accessories with the IH 58 to ensure correct fit and performance; the manual advises against modifications or non-approved parts. Some third-party mouthpieces or masks may fit, but using non-certified parts risks reduced nebulisation efficiency or hygiene issues. For best results and warranty compliance, purchase Beurer replacement atomisers, filters, and masks made for the IH 58.

How do I clean and disinfect the beurer Nebuliser IH 58 to prevent common problems?

Clean the atomiser, mouthpiece, and mask after each use with hot, non-boiling water and let them air dry; do not immerse the compressor or tubing in water. Perform daily disinfection of accessories (per manual recommendations) using a recommended disinfectant or boiling method if allowed for the specific parts. Regular cleaning prevents blockages, bacterial growth, and common problems like weak mist output—replace components that show wear or discoloration.

More beurer manuals